Short communication: Automated testing of accelerometer transverse sensitivity with a flexure-based XY compliant stage

Tiemin Li, Yunsong Du, Wei Ji, Zhihua Liu, and Chenguang Cai

Viewed

Total article views: 1,213 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
700 459 54 1,213 52 51
  • HTML: 700
  • PDF: 459
  • XML: 54
  • Total: 1,213
  • BibTeX: 52
  • EndNote: 51
Views and downloads (calculated since 02 Aug 2016)
Cumulative views and downloads (calculated since 02 Aug 2016)

Cited

Latest update: 28 Mar 2024
Download
Short summary
This paper presents the automated testing of accelerometer transverse sensitivity with a flexure-based compliant stage. An experimental platform is set up, and the circular tracking error is first obtained to verify the feasibility of the stage. The transverse sensitivity of the accelerometer under test is below 3 %. The difference of the transverse sensitivity under various frequencies is 0.4 %, while the difference of measured direction angle is 2.49, which validates the accuracy of the method.